Quality Inspection Report of DXWD-80 Lab Electromagnet

Home    Dexinmag News    Electromagnet    Quality Inspection Report of DXWD-80 Lab Electromagnet

 

DXWD-80 Electromagnet  DXWD-80 Electromagnet  DXWD-80 Electromagnet

Overall Dimensions: 549mm(L) x 313mm(W) x 335mm(H)

Coil spacing: 80mm
Pole diameter: 80mm
Pole surface diameter: 50mm

Air gap range: 0-80mm

Weight: 119.4kg

 

Performance test report
 Introduction
    1. Test purpose
Through the analysis of the test results, the evaluation of the product performance is obtained;
Evaluate whether the test product meets the technical requirements;
Provide reference data for product design and after-sales service, and provide a strong basis for leaving the factory.
    2. Basic information
 Product Name:
Electromagnet
Product model: DXWD-80
Test time: June 06, 2023
    3. Testing environment
The detection environment is closer to the actual environmental conditions used by the user, including factors such as lead wire length and ambient temperature. The time is selected in the time interval where the fluctuation of the geomagnetic field is small. Before the test, the instruments required for the test were turned on and warmed up in advance and the measurement environment was recorded, as shown in Table 1.
Measurement environment
Environment temperature Product initial temperature Lead length Environmental magnetic field
20.0℃ 20.0℃ 4m 0~30000nT
Table 1 Measurement environment data
 
Magnetic field parameter test
    1. Using the device
DX-F2031 power supply
DX-180 Gauss Meter
    2. Magnetic field measurement
This product uses the following measurement methods:
Fixed-point measurement: measure the magnetic field at the target point
Magnetic field strength
Air Gap Display current (A) Display voltage Voltage (V) Magnetic field
strength(Gs)
20mmGap 5 - 0.403
20mmGap 10 - 0.799
20mmGap 13 - 1.013
20mmGap 15 - 1.133
Table 2 Magnetic field strength data
 
Electrical parameter test
    1. Using equipment
Digital multimeter
Impedance analyzer
    2. Test method
The digital instrument with bridge circuit and voltammetry is used for measurement. The measurement results are shown in the table below.
Resistance R measurement
Environment
temperature
Product temperature Measuring resistance ResistanceR@20℃ ResistanceR@85℃
20.0℃ 20.0℃ 2.280Ω 2.280Ω 2.962Ω
Table 3 Resistance R data
 
Conclusion
The product's magnetic field and electrical performance meet the requirements of the corresponding technical parameters. The final performance of the product is subject to this document.

 

 

Related Products

 

Electromagnet      Helmholtz Coil      Solenoid      Constant Magnetic Field Permanent Magnet