Quality Inspection Report: DX0410-005 Calibration Magnet
Overall Dimensions: 82mm(L) x 82mm(W) x 102mm(H)
Polar diameter: 40mm
Air gap: 10mm
Weight: 2.217kg
Performance Testing Report | |||||||
1. Summary | |||||||
1)Test Objectives | |||||||
Through analysis of the test results, obtain an evaluation of the product's performance; Assess whether the tested product meets technical requirements; Provide reference data for product design and after-sales service, and provide a strong basis for factory delivery. |
|||||||
2)Basic information | |||||||
NO: | 117 | Model: | DX0410-005 | ||||
Item name: | Calibration magnet | testing time: | 20240403 | ||||
3)Testing environment | |||||||
The testing environment is selected to closely resemble the actual operating conditions of the user, including factors such as lead length and ambient temperature. The time chosen should be within a period of minimal fluctuation in the geomagnetic field. Prior to testing, the necessary instruments should be powered on and allowed to warm up, and the measurement environment should be recorded, as shown in Table 1. | |||||||
Testing environment | |||||||
T | FT | Wires length | E magnetic field | ||||
11.0℃ | 11.0℃ | - | 0~30000nT | ||||
Table Testing environment data | |||||||
2. Magnetic Field parameter testing | |||||||
1)Equipment | |||||||
DX-180 Gaussmeter | |||||||
2)Magnetic field measurement | |||||||
This product employs the following measurement methods: Fixed Point Measurement: Measurement of the magnetic field at specific points. |
|||||||
Magnetic field | |||||||
Air Gap | A | V | Magnetic field intensity | ||||
10mmGap | - | - | 510.58Gs | ||||
Table 2 Magnetic field testing data | |||||||
Uniformity | |||||||
NO | Area | A | Max | Min | Uniformity | Note | |
1 | φ6*6mm | - | 510.98Gs | 510.07Gs | 0.10% | ||
Table 3 Uniformity area | |||||||
3. Conclusion | |||||||
The magnetic field and electrical performance of this product both meet the corresponding technical parameter requirements. The final performance of the product is based on this document. |
You may also like
Calibration Certificate of DX0410-005 Calibration magnet